FDA Adverse Event
Malfunction
Summary report: N
STEALTHSTATION TRIA NAVIGATION SYSTEM
MDR report key: 3002367
·
Received March 13, 2013
Report
- Report Number
- 1723170-2013-00173
- Event Type
- Malfunction
- Date Received
- March 13, 2013
- Date of Event
- May 6, 2013
- Report Date
- May 6, 2013
- Manufacturer
- MEDTRONIC NAVIGATION, INC.
- Product Code
- HAW
- PMA / PMN Number
- K001153
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
NO PATIENT WAS PRESENT AT THE TIME OF THE ALLEGED MALFUNCTION. NO PARTS OR FILES HAVE BEEN RECEIVED BY THE MANUFACTURER FOR EVALUATION.
Additional Manufacturer Narrative · 1
SUSPECT COMPUTER EVALUATION FINDS: TAMPER SEALS BROKEN. POWERED UP SYSTEM, BOOTED NORMALLY, BUT WITH AUDIBLE FAN NOISE. INTERNAL INSPECTION REVEALED THAT THE VIDEO GRAPHICS CARD (VGC) COOLING FAN IS NOT AT FULL OPERATION. VGC COULD OVERHEAT AND CAUSE REPORTED ISSUE. GRAPHICS CARD MALFUNCTION DIRECTLY CAUSED EVENT. REPLACEMENT COMPUTER SENT TO SITE AND INSTALLED IN SYSTEM FOR ISSUE RESOLUTION.
Description of Event or Problem · 1
A MEDTRONIC REPRESENTATIVE REPORTED A STEALTHSTATION TRIA NAVIGATION SYSTEM COMPUTER THAT BECAME UNRESPONSIVE. THERE WAS NO PATIENT PRESENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 105442 | STEALTHSTATION TRIA NAVIGATION SYSTEM | NEUROLOGICAL STEREOTAXIC INSTRUMENT | HAW | MEDTRONIC NAVIGATION, INC. | TRIA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |