FDA Adverse Event Malfunction Summary report: N

LINOX S 65

MDR report key: 3002323 · Received March 13, 2013

Report

Report Number
1028232-2013-00711
Event Type
Malfunction
Date Received
March 13, 2013
Date of Event
February 14, 2013
Report Date
March 4, 2013
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVY
PMA / PMN Number
P980023
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT, THE LEAD WAS FOUND DISSECTED SOME 25 CM DISTAL TO THE IS-1 CONNECTOR PIN. BOTH FRAGMENTS WERE RETURNED FOR ANALYSIS. THE ANALYSIS OF THE LEAD DEMONSTRATED A RUBBED THROUGH INSULATION. THIS INSULATION DAMAGE CAN BE CONSIDERED AS THE ROOT CAUSE OF THE ISSUES MENTIONED IN THE COMPLAINT DESCRIPTION. BASED ON THE CHARACTERISTICS OF THIS DAMAGE, IT IS REASONABLE TO ASSUME THAT THE LEAD HAD BEEN SUBJECT TO SEVERE MECHANICAL STRESS IN THE IMPLANTED STATE. DIAGNOSTIC IMAGES CLARIFYING THIS ASSUMPTION WERE NOT AVAILABLE FOR ANALYSIS. THE ANALYSIS DID NOT REVEAL ANY SIGN OF A MATERIAL OR MANUFACTURING PROBLEM.

Description of Event or Problem · 1

THIS LEAD WAS EXPLANTED AND REPLACED DUE TO NOISE, INTERMITTENT OVERSENSING, AND LOW IMPEDANCE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
104691 LINOX S 65 ICD LEAD NVY BIOTRONIK SE & CO. KG 351333

Patients

Seq Age Sex Outcome Treatment
1 72 YR Hospitalization