FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3002289 · Received March 13, 2013

Report

Report Number
2531779-2013-02674
Event Type
Malfunction
Date Received
March 13, 2013
Report Date
March 4, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 (B)(4) - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE TOTAL DAILY DOSE HISTORY INDICATED THAT INSULIN DELIVERY TOTALS CORRECTLY REFLECTED PROGRAMMED VALUES. THERE WERE NO ALARMS OR WARNINGS RELATED TO THE COMPLAINT OBSERVED IN THE BLACK BOX OR ALARM HISTORY; ONLY TYPICAL USAGE WAS OBSERVED. THE PUMP WAS EXERCISED FOR 29 HOURS WITH NO DELIVERY ISSUES. THE PUMP WAS EVALUATED AND FOUND TO BE OPERATING WITHIN REQUIRED SPECIFICATIONS AND DELIVERING INSULIN ACCURATELY.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

THE REPORTER CONTACTED ANIMAS ON (B)(6) 2013 REPORTING THAT THE PATIENT'S BLOOD GLUCOSE (BG) WAS RUNNING AS HIGH AS 20MMOL/L. THE REPORTED BG WAS NOT INDICATIVE OF A SERIOUS INJURY AND DOES NOT MEET THE ANIMAS CRITERIA OF AN ADVERSE EVENT. THE REPORTER STATED THE PATIENT HAS BEEN COVERING BG MORE FREQUENTLY AND WITH GREATER AMOUNTS THAN NORMAL. THE MOTOR SOUNDS DIFFERENT THAN OTHER PUMP THAT IT IS LOUD AND GRINDING. THE HISTORY AND SETTINGS ALL CONFIRMED TO BE ACCURATE. THIS REPORT IS BEING MADE DUE TO THE INACCURATE DELIVERY ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
104648 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 9 YR