FDA Adverse Event Malfunction Summary report: N

ETRIO

MDR report key: 3002219 · Received March 13, 2013

Report

Report Number
3005075853-2013-01233
Event Type
Malfunction
Date Received
March 13, 2013
Date of Event
February 22, 2013
Report Date
February 28, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GEI
PMA / PMN Number
K072177
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). SHOULD THE INFORMATION BE PROVIDED LATER, A SUPPLEMENTAL MEDWATCH WILL BE SENT. THE DEVICE WAS RECEIVED WITH THE ELECTRODE SEPARATED FROM THE CERAMIC AND THE ELECTRODE IS STILL ATTACHED TO THE ACTIVE ROD. TESTING FOUND A SHORT ON THE DEVICE WHEN THE JAWS ARE FULLY OR PARTIALLY CLOSED, CAUSING A REPLACE LIGHT TO ILLUMINATE. THE SEPARATED ELECTRODE ALLOWS THE ELECTRODE TO BE TOUCHING THE JAWS WHEN THEY ARE CLOSED. THE ELECTRODE TO JAW CONTACT CREATES AN ELECTRICAL SHORT PREVENTING RF POWER DELIVERY FROM THE GENERATOR RESULTING IN THE REPLACE LIGHT ILLUMINATING ON THE RF60.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC HYSTERECTOMY PROCEDURE THE SURGEON WAS USING THE DEVICE TO DISSECT AND SEAL A VESSEL WHEN HE NOTICED THAT A FILAMENT WAS DISPLACED FROM THE JAW OF THE DEVICE. AS THIS HAS HAPPENED BEFORE, HE STOPPED USING THE DEVICE AND REMOVED IT FROM THE PATIENT AND COMPLETED THE CASE WITH ANOTHER DEVICE. NO PATIENT CONSEQUENCES WERE REPORTED. ONE DEVICE WILL BE RETURNING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
105504 ETRIO ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES GEI ETHICON ENDO-SURGERY, LLC. UNK J4CR4C

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR AND HANDPIECE