FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3002205
·
Received March 13, 2013
Report
- Report Number
- 2032227-2013-00961
- Event Type
- Malfunction
- Date Received
- March 13, 2013
- Date of Event
- January 23, 2013
- Report Date
- February 28, 2013
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IS
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP ALARM MOTOR ERROR DURING THE BASIC OCCLUSION TEST, DUE TO LOOSE DRIVE SUPPORT DISK. UNIT WAS MISSING END CAP STICKER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED DURING MANUAL PRIME. NOTHING FURTHER WAS REPORTED.