FDA Adverse Event Malfunction Summary report: N

HEARTSINE SAMARITAN 300P AND PAD PAK

MDR report key: 3002175 · Received February 11, 2013

Report

Report Number
3004123209-2013-00392
Event Type
Malfunction
Date Received
February 11, 2013
Date of Event
February 5, 2012
Report Date
January 21, 2013
Manufacturer
HEARTSINE TECHNOLOGIES, LTD
Product Code
MKJ
PMA / PMN Number
K052465
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE PAD DEVICE WAS INSTALLED ON (B)(6) 2009 AND OPERATED SUCCESSFULLY UNTIL (B)(6) 2012. ON (B)(6) 2012 THE DEVICE FAILED A WEEKLY SELF TEST DUE TO A LOW BATTERY. THE DEVICE REMAINED IN FAULT MODE UNTIL (B)(6) 2012. A NEW PAD-PAK WAS INSERTED BUT THE DEVICE STARTS TO SWITCH ITSELF ON AUTOMATICALLY. THE PROBLEM WITH THE DEVICE WAS ATTRIBUTED TO A FAULT IN THE MEMBRANE. THE DEVICE FAILED A SELF TEST DUE TO A LOW BATTERY WHICH TRIGGERED THE BATTERY MANAGEMENT SYSTEM. THE LOW BATTERY WAS ALMOST CERTAINLY CAUSED BY THE EXPOSURE TO VERY LOW TEMPERATURES. THE PAD PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELLED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.

Description of Event or Problem · 1

THERE WAS NO PT INVOLVED IN THIS EVENT. THIS DEVICE MALFUNCTIONED BECAUSE IT WAS SWITCHING ITSELF ON AUTOMATICALLY. A DEVICE SWITCHING ITSELF ON AUTOMATICALLY, IF LEFT UNDETECTED COULD RESULT IN THE BATTERY BECOMING DEPLETED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
60130 HEARTSINE SAMARITAN 300P AND PAD PAK AUTOMATED EXTERNAL DEFIBRILLATOR MKJ HEARTSINE TECHNOLOGIES, LTD PAD

Patients

Seq Age Sex Outcome Treatment
1