FDA Adverse Event
Malfunction
Summary report: N
HEARTSINE SAMARITAN 300P AND PAD PAK
MDR report key: 3002171
·
Received February 11, 2013
Report
- Report Number
- 3004123209-2013-00157
- Event Type
- Malfunction
- Date Received
- February 11, 2013
- Date of Event
- January 4, 2013
- Report Date
- February 6, 2013
- Manufacturer
- HEARTSINE TECHNOLOGIES, LTD
- Product Code
- MKJ
- PMA / PMN Number
- K052465
- Removal / Correction Number
- Z-0124-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
ROUTINE TESTING CONFIRMED THAT THIS DEVICE WAS INSTALLED BY THE CUSTOMER IN (B)(6) 2010 AND THAT IT HAD PERFORMED TO SPEC UP TO (B)(6) 2012. INVESTIGATION CONFIRMED THE DEVICE FAILED TO UPGRADE TO SOFTWARE VERSION 3.2.0 USING THE HEARTSINE SAMARITAN PAD UNIVERSAL UPGRADE. TESTING ALSO CONFIRMED THAT THE DEVICE WAS CAPABLE OF DELIVERING THERAPY IN THIS FAULT MODE. THIS DEVICE WAS REPLACED WITHIN 24 HOURS. THE PAD PAK, WHICH CONTAINS THE ELECTRODE AND BATTERIES, IS LABELLED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.
Description of Event or Problem · 1
THERE WAS NO PT INVOLVED IN THIS EVENT. THE DEVICE MALFUNCTIONED THAT THE SOFTWARE FAILED TO UPGRADE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 60176 | HEARTSINE SAMARITAN 300P AND PAD PAK | AUTOMATED EXTERNAL DEFIBRILLATOR | MKJ | HEARTSINE TECHNOLOGIES, LTD | PAD |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |