FDA Adverse Event Malfunction Summary report: N

HEARTSINE SAMARITAN 300P AND PAD PAK

MDR report key: 3002171 · Received February 11, 2013

Report

Report Number
3004123209-2013-00157
Event Type
Malfunction
Date Received
February 11, 2013
Date of Event
January 4, 2013
Report Date
February 6, 2013
Manufacturer
HEARTSINE TECHNOLOGIES, LTD
Product Code
MKJ
PMA / PMN Number
K052465
Removal / Correction Number
Z-0124-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

ROUTINE TESTING CONFIRMED THAT THIS DEVICE WAS INSTALLED BY THE CUSTOMER IN (B)(6) 2010 AND THAT IT HAD PERFORMED TO SPEC UP TO (B)(6) 2012. INVESTIGATION CONFIRMED THE DEVICE FAILED TO UPGRADE TO SOFTWARE VERSION 3.2.0 USING THE HEARTSINE SAMARITAN PAD UNIVERSAL UPGRADE. TESTING ALSO CONFIRMED THAT THE DEVICE WAS CAPABLE OF DELIVERING THERAPY IN THIS FAULT MODE. THIS DEVICE WAS REPLACED WITHIN 24 HOURS. THE PAD PAK, WHICH CONTAINS THE ELECTRODE AND BATTERIES, IS LABELLED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.

Description of Event or Problem · 1

THERE WAS NO PT INVOLVED IN THIS EVENT. THE DEVICE MALFUNCTIONED THAT THE SOFTWARE FAILED TO UPGRADE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
60176 HEARTSINE SAMARITAN 300P AND PAD PAK AUTOMATED EXTERNAL DEFIBRILLATOR MKJ HEARTSINE TECHNOLOGIES, LTD PAD

Patients

Seq Age Sex Outcome Treatment
1