HEARTSINE SAMARITAN 300P AND PAD PAK
Report
- Report Number
- 3004123209-2013-00470
- Event Type
- Malfunction
- Date Received
- February 11, 2013
- Date of Event
- June 4, 2011
- Report Date
- January 21, 2013
- Manufacturer
- HEARTSINE TECHNOLOGIES, LTD
- Product Code
- MKJ
- PMA / PMN Number
- K052465
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE PAD DEVICE WAS INSTALLED ON (B)(6) 2011 AND OPERATED SUCCESSFULLY UNTIL (B)(6) 2011. BETWEEN (B)(6) 2011 THERE ARE 22 MANUAL SELF TESTS RECORDED. ON (B)(6) 2011 THE DEVICE RECORDS A MANUAL SELF TEST DUE TO A LOW BATTERY. MULTIPLE MANUAL POWER UPS WOULD INDICATE THE DEVICE WAS SWITCHING ITSELF ON AUTOMATICALLY. THE PROBLEM WITH THE DEVICE WAS ATTRIBUTED TO A FAULT IN THE MEMBRANE. THERE IS ALSO EVIDENCE OF CORROSION ON THE DEVICE MEMBRANE. THIS WOULD INDICATE THE DEVICE WAS BEING EXPOSED TO ADVERSE ENVIRONMENTAL CONDITIONS WHICH CAN HAVE A DETRIMENTAL EFFECT ON THE DEVICE MEMBRANE. THE PAD PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELLED FOR SINGLE USE BUT NOT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.
THERE WAS NO PATIENT INVOLVED IN THIS EVENT. THIS DEVICE MALFUNCTIONED BECAUSE IT WAS SWITCHING ITSELF ON AUTOMATICALLY. A DEVICE SWITCHING ITSELF ON AUTOMATICALLY, IF LEFT UNDETECTED COULD RESULT IN THE BATTERY BECOMING DEPLETED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 59245 | HEARTSINE SAMARITAN 300P AND PAD PAK | AUTOMATED EXTERNAL DEFIBRILLATOR | MKJ | HEARTSINE TECHNOLOGIES, LTD | PAD |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |