FDA Adverse Event Malfunction Summary report: N

HEARTSINE SAMARITAN 300P AND PAD PAK

MDR report key: 3002109 · Received February 11, 2013

Report

Report Number
3004123209-2013-00493
Event Type
Malfunction
Date Received
February 11, 2013
Date of Event
October 16, 2011
Report Date
January 21, 2013
Manufacturer
HEARTSINE TECHNOLOGIES, LTD
Product Code
MKJ
PMA / PMN Number
K052465
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE PAD DEVICE WAS INSTALLED ON (B)(6) 2010 AND OPERATED SUCCESSFULLY UNTIL (B)(6) 2011. ON (B)(6) 2011, THE DEVICE FAILED A WEEKLY SELF TEST FOR A LOW BATTERY. A NEW PAD-PAK WAS INSTALLED THAT DAY AND THE DEVICE OPERATED UNTIL (B)(6) 2012. ON (B)(6) 2012, THERE WERE MULTIPLE MANUAL POWER UPS, WHICH WOULD INDICATE THE DEVICE WAS SWITCHING ITSELF ON AUTOMATICALLY AND THE PAD-PAK WAS REMOVED. A NEW PAD-PAK WAS AGAIN INSTALLED ON (B)(6) 2012, AND THE DEVICE PASSED AN AUTO SELF-TEST. THE PROBLEM WITH THE DEVICE WAS ATTRIBUTED TO A FAULT IN THE MEMBRANE. THE PAD-PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELLED FOR SINGLE USE, BIT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.

Description of Event or Problem · 1

THERE WAS NO PT INVOLVED IN THIS EVENT. THIS DEVICE MALFUNCTIONED BECAUSE IT WAS SWITCHING ITSELF ON AUTOMATICALLY A DEVICE SWITCHING ITSELF ON AUTOMATICALLY, IF LEFT UNDETECTED, COULD RESULT IN THE BATTERY BECOMING DEPLETED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
60174 HEARTSINE SAMARITAN 300P AND PAD PAK AUTOMATED EXTERNAL DEFIBRILLATOR MKJ HEARTSINE TECHNOLOGIES, LTD PAD