FDA Adverse Event Malfunction Summary report: N

ITREL 4

MDR report key: 3002080 · Received March 13, 2013

Report

Report Number
3004209178-2013-03643
Event Type
Malfunction
Date Received
March 13, 2013
Date of Event
February 19, 2013
Report Date
April 2, 2013
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LGW
PMA / PMN Number
P840001
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

CONCOMITANT PRODUCTS: PRODUCT ID 7495LZ51, SERIAL# (B)(4), IMPLANTED: (B)(6) 2003, PRODUCT TYPE EXTENSION; PRODUCT ID 7434A, SERIAL# (B)(4), IMPLANTED: (B)(6) 2004, PRODUCT TYPE PROGRAMMER, PATIENT; PRODUCT ID 3888-28, LOT# J0225319V, IMPLANTED: (B)(6) 2003, PRODUCT TYPE LEAD. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT AFTER THE PATIENT'S DEVICE WAS REPLACED, THE PATIENT WAS RECEIVING GOOD STIMULATION AND GOOD THERAPY. PLEASE REFER TO MFR. REPORT NO. 6000032-2013-00059 REGARDING PATIENT'S PREVIOUS DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WERE HIGH IMPEDANCES AND A LOSS OF STIMULATION FOLLOWING AN IMPLANTABLE NEUROSTIMULATOR (INS) REPLACEMENT SURGERY. IT WAS STATED THAT THE IMPEDANCE READINGS WERE GREATER THAN 40,000 OHMS MEASURED AT THREE VOLTS, AND IT WAS AFFECTING THE CASE PAIRS ONLY. IT WAS ALSO STATED THAT ALL UNIPOLAR COMBINATIONS WERE GREATER THAN 40,000. THE FOLLOWING BIPOLAR IMPEDANCES WERE REPORTED: '01 =1400, 02=2300, 03=3300, 12=1400, 13=1300, 23=2300' (THE CALLER ROUNDED THE NUMBERS). THE CALLER STATED HE SAW 'GREATER THAN >4,000 IMPEDANCES ON PREVIOUS INS, HOWEVER, CALLER THEN REPORTED PATIENT LOST THERAPY APPROXIMATELY THREE WEEKS PRIOR TO REPORT. ADDITIONAL INFORMATION HAS BEEN REQUESTED, AND IF RECEIVED, A SUPPLEMENTAL REPORT WILL BE FILED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
104677 ITREL 4 STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF LGW MEDTRONIC MED REL MEDTRONIC PUERTO RICO 37703

Patients

Seq Age Sex Outcome Treatment
1 63 YR