FDA Adverse Event Injury Summary report: N

SETROX S 53

MDR report key: 3002042 · Received March 13, 2013

Report

Report Number
1028232-2013-00660
Event Type
Injury
Date Received
March 13, 2013
Date of Event
January 23, 2013
Report Date
February 28, 2013
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVN
PMA / PMN Number
P950037
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ASSOCIATED WITH THE MANUFACTURE OF THIS PARTICULAR DEVICE. THE MANUFACTURING PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MANUFACTURING PROCESS WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MANUFACTURING.

Additional Manufacturer Narrative · 1

THE ANALYSIS OF THE LEAD DEMONSTRATED A DAMAGED INSULATION AND A FRACTURED OUTER COIL AT A DISTANCE OF SOME 24 CM DISTAL TO THE IS-1 CONNECTOR PIN. IN THAT SECTION, THE LEAD BODY WAS FOUND SQUEEZED AND DEFORMED. BASED ON THE CHARACTERISTICS AS WELL AS THE LOCATION OF THE DAMAGE, IT IS REASONABLE TO ASSUME THAT THE LEAD HAD BEEN SUBJECT TO EXCESSIVE MECHANICAL FORCES AS THE RESULT OF CLAVICULAR - FIRST RIB ENTRAPMENT. THE ANALYSIS DID NOT REVEAL ANY SIGN OF A MATERIAL OR MANUFACTURING PROBLEM.

Description of Event or Problem · 1

THIS LEAD WAS EXPLANTED AND REPLACED ON (B)(6) 2013 DUE TO SUBCLAVIAN CRUSH AND UNDER-SENSING. THIS OCCURRED ABOUT 2 WEEKS AFTER THE PATIENT UNDERWENT AORTIC VALVE SURGERY. THE HOSPITAL RETAINED THE DEVICE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
104635 SETROX S 53 PACER LEAD NVN BIOTRONIK SE & CO. KG 350974

Patients

Seq Age Sex Outcome Treatment
1 68 YR Hospitalization