FDA Adverse Event Malfunction Summary report: N

DEXTRUS 4136

MDR report key: 3002023 · Received March 13, 2013

Report

Report Number
1028232-2013-00609
Event Type
Malfunction
Date Received
March 13, 2013
Date of Event
January 17, 2013
Report Date
February 27, 2013
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVN
PMA / PMN Number
P950037
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ASSOCIATED WITH THE MANUFACTURE OF THIS PARTICULAR DEVICE. THE MANUFACTURING PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MANUFACTURING PROCESS WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MANUFACTURING.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT ATRIAL LEAD EXHIBITED NOISE WHICH LEAD TO INAPPROPRIATE MODE SWITCHES. A LEAD SAFETY SWITCH WAS NOTED WHICH CHANGED THE LEAD CONFIGURATION FROM BIPOLAR TO UNIPOLAR. A REVIEW OF THE DAILY LOGBOOK NOTED THAT THERE WERE 2 DAYS WHICH SHOWED THE LEAD IMPEDANCE AT N/R. ISOMETRICS WERE PERFORMED WHICH FOUND NOISE AND OVERSENSING, ALL OTHER MEASUREMENTS WERE WITHIN NORMAL RANGES. BOSTON SCIENTIFIC TECHNICAL SERVICES WAS CONSULTED AND SUGGESTED DOING A CHEST X-RAY TO DETERMINE IF THE LEAD HAD MOVED AT ALL. AT THIS TIME, THE LEAD REMAINS IMPLANTED. TO DATE, NO ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
105791 DEXTRUS 4136 PACER LEAD NVN BIOTRONIK SE & CO. KG 358754

Patients

Seq Age Sex Outcome Treatment
1 Other