FDA Adverse Event Malfunction Summary report: N

AXON ECLIPSE CONTROLLER

MDR report key: 3001919 · Received February 12, 2013

Report

Report Number
1045254-2013-00091
Event Type
Malfunction
Date Received
February 12, 2013
Report Date
November 27, 2012
Manufacturer
XOMED MFG JACKSONVILLE
Product Code
GWF
PMA / PMN Number
K061639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

TESTING/REPAIR RECORDS CONFIRMED THE REPORTED EVENT AND VERIFIED THE ERROR MESSAGE. THE STIMULATION BOARD AND CABLES WERE REPLACED. THE MFG DATE IS UNOBTAINABLE. RECORDS PRIOR TO SEPTEMBER 2011 REMAIN WITH THE ORIGINAL DEVICE MFR. THE NIM-ECLIPSE SYSTEM NEUROVASCULAR WORKSTATION IS INTENDED FOR USE TO MONITOR SENSORY AND MOTOR PATHWAYS. IF THE SYSTEM FAILS TO PERFORM AS INTENDED, (ESPECIALLY TO EFFECT OR DETECT ELECTROMYOGRAPHIC (EMG) ACTIVITY) IT PRESENTS THE POTENTIAL FOR TEMPORARY OR PERMANENT DAMAGE TO THE NERVE THAT IS PRESENT. THE ECLC ECLIPSE SYSTEM CONTROLLER PROVIDES HIGH SPEED DIGITAL DATA PROCESSING, STIMULATION GENERATION AND AUDIO PROCESSING OF EMG ACTIVITY AND ALSO SUPPLIES SWITCHED AC POWER TO THE COMPUTER. THE CONTROLLER CONNECTS TO THE COMPUTER VIA THE HIGH SPEED PCMCIA INTERFACE. WHEN INFO SUGGESTS THAT THE NIM ECLIPSE EQUIPMENT HAD THE POTENTIAL TO NOT STIMULATE TO AFFECT ELECTROMYOGRAPHY (EMG), OR TO DETECT EMG, IT IS ASSUMED THAT THE FAILURE WAS DEVICE-RELATED AND MAY HAVE GONE UNDETECTED. IN THIS CASE, THERE IS NO INFO TO SUGGEST AN EVENT COULD NOT BE INTERPRETED FALSELY - THE REPORT IS INCONCLUSIVE AS TO THE CAUSE OF FIELD OBSERVATION. THEREFORE, W/O INFO TO REASONABLY SUGGEST SERIOUS INJURY, OR THAT MEDICAL INTERVENTION WAS REQUIRED TO PRECLUDE SERIOUS INJURY, WE ARE FILING THIS REPORT AS A PRODUCT PROBLEM. THIS PRODUCT WAS BEING USED FOR TREATMENT, NOT DIAGNOSIS.

Description of Event or Problem · 1

THE MFR HAS REFINED THE CRITERIA FOR MAKING MDR DECISIONS. UPON A RETROSPECTIVE REVIEW, THE FOLLOWING EVENT IS NOT BELIEVED TO BE REPORTABLE: A CUSTOMER RETURNED AN ECLIPSE CONTROLLER STATING "THAT ITS DUE FOR IT'S ANNUAL CHECKUP/PREVENTATIVE MAINTENANCE. IN ADDITION, THE CONTROLLER RECENTLY HAD AN ERROR RELATING TO THE COOLING SYSTEM NOT WORKING PROPERLY." THERE WAS NO SUGGESTION OF PT INJURY OR INVOLVEMENT. TESTING/REPAIR DIE NOT CONFIRM THE REPORTED EVENT, BUT DID FIND A FAILED STIMULATION BOARD. A FAILED STIMULATION BOARD HAS THE POTENTIAL TO CAUSE OR CONTRIBUTE TO PT INJURY (LOSS OF STIMULATION DURING USE LEADING TO A FALSE NEGATIVE) - AS A SUDDEN FAILURE IS NOT CONSISTENTLY ACCOMPANIED BY WARNING/ERROR MESSAGES DURING A PROCEDURE. THE ERROR MESSAGE CITED IS NOT RELATED TO THE FAILED BOARD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
61984 AXON ECLIPSE CONTROLLER GWF - STIMULATOR, ELECTRICAL, EVOKED RESPONSE GWF XOMED MFG JACKSONVILLE 945ECLC NI

Patients

Seq Age Sex Outcome Treatment
1