FDA Adverse Event Malfunction Summary report: N

7900

MDR report key: 3001885 · Received February 12, 2013

Report

Report Number
9680959-2013-00353
Event Type
Malfunction
Date Received
February 12, 2013
Date of Event
January 15, 2013
Report Date
February 12, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS GMBH
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REP EVALUATED THE SYS AND REPLACED THE CPU. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYSTEM WOULD NOT COMPLETE BOOT UP. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
61910 7900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS GMBH 7900

Patients

Seq Age Sex Outcome Treatment
1