FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3001843 · Received February 12, 2013

Report

Report Number
1720753-2013-01684
Event Type
Malfunction
Date Received
February 12, 2013
Date of Event
January 31, 2013
Report Date
February 12, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE POWER SUPPLY WAS ADJUSTED AND THE HIGH VOLTAGE CABLE WAS CLEANED AND REGREASED DURING THE SERVICE CALL. THE REPORTED ISSUE IS CURRENTLY UNDER INVESTIGATION.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A GENERATOR ERROR AND A CONTROLLER ERROR AND THE SYSTEM AUTOMATICALLY SHUTS DOWN WHEN THIS OCCURS. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
61356 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1