FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 3001685
·
Received March 8, 2013
Report
- Report Number
- 8020893-2013-00563
- Event Type
- Malfunction
- Date Received
- March 8, 2013
- Date of Event
- January 30, 2013
- Report Date
- February 7, 2013
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT 840 VENTILATOR HAD A BLANK SCREEN. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE GRAPHIC USER INTERFACE CPU PCB AND THE LCD FLEX CIRCUIT. THE UNIT PASSED EXTENDED SELF-TESTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 99871 | 840 VENTILATOR | VENTILATOR | CBK | COVIDIEN, FORMERLY NELLCOR | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |