FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 3001497 · Received March 5, 2013

Report

Report Number
1627487-2013-12306
Event Type
Injury
Date Received
March 5, 2013
Date of Event
February 13, 2013
Report Date
February 13, 2013
Manufacturer
ST JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS CHARGER MODEL NUMBER WAS INCLUDED IN A FIELD CORRECTION. CORRECTIVE AND PREVENTIVE ACTION (CAPA). RESULTS - POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGES WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFO RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2. REF MFR REPORT #1627487-2013-12305. IT WAS REPORTED THE PT EXPERIENCES HEATING AT THE IPG POCKET SITE WHILE CHARGING. A REPLACEMENT CHARGER WAS SENT TO THE PT. ON (B)(4) 2012 ST JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
93071 CHARGING SYSTEM SCS CHARGER LGW ST JUDE MEDICAL - NEUROMODULATION 3721 3694019

Patients

Seq Age Sex Outcome Treatment
1 60 YR Required Intervention IMPLANT DATE:| SCS LEAD: MODEL 3163| IMPLANT DATE:| SCS LEAD: MODEL 3186