CHARGING SYSTEM
Report
- Report Number
- 1627487-2013-05317
- Event Type
- Injury
- Date Received
- March 5, 2013
- Date of Event
- February 11, 2013
- Report Date
- February 11, 2013
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. CORRECTIVE AND PREVENTIVE ACTION (CAPA) WAS PERFORMED. POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 2 OF 2. REF MFR REPORT #: 1627487-2013-05316. IT WAS REPORTED THE PT HAS BEEN EXPERIENCING HEATING WHILE RECHARGING THE IPG. THE PT ATTEMPTED TO RECHARGE IN SHORT INTERVALS TO NO AVAIL. THE PT WAS SENT A REPLACEMENT CHARGING SYSTEM AND IT PERFORMS BETTER THAN THE PRIOR CHARGING SYSTEM. THE PT'S SCS SYSTEM WILL BE EXPLANTED DUE TO THE PT NEEDING AN MRI. THE PT WILL MEET WITH HIS DOCTOR ON A LATER DATE. ON 08/01/2012, ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 93991 | CHARGING SYSTEM | SCS CHARGIN SYSTEM | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3721 | 3126004 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | Required Intervention | SCS ANCHOR, MODEL 1192| IMPLANT DATE:| SCS LEAD, MODEL 3186| IMPLANT DATE:| IMPLANT DATE:| SCS LEAD, MODEL 3186 |