FDA Adverse Event Malfunction Summary report: N

IMPAX CV REPORTING

MDR report key: 3001215 · Received February 26, 2013

Report

Report Number
1225058-2013-00042
Event Type
Malfunction
Date Received
February 26, 2013
Date of Event
January 29, 2013
Report Date
February 26, 2013
Manufacturer
AGFA HEALTHCARE CORP.
Product Code
LLZ
PMA / PMN Number
K050228
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AGFA'S INVESTIGATION OF THIS MDR REVEALED THAT IN 2012, A CORRECTION WAS IMPLEMENTED BY A AGFA CLINICAL ANALYST RECONFIGURING THE CUSTOMER'S SYSTEM TO ELIMINATE THE TYPE OF MISMAPPING SEEN IN THIS CURRENT MDR (THE 2012 EVENT AND CORRECTION WERE REPORTED VIA FDA MDR 1225058-2012-00002). THIS PAST CORRECTION WAS MISTAKENLY REVERTED TO IT'S PREVIOUS CONFIGURATION IN IMPAX CV FOR PHILIPS IE-33 ULTRASOUND CARTS BY A AGFA PROFESSIONAL SERVICES PERSON, THUS RE-ENABLING THE ERRORS INITIALLY FOUND ON (B)(6) 2012. ON (B)(4) 2013, AGFA CORRECTED THE ISSUE BY RECONFIGURING THE CUSTOMER'S SYSTEM AND ELIMINATING THE MISMAPPING. THE CUSTOMER HAS CONFIRMED THE IMPLEMENTED CORRECTION HAS FIXED THE PROBLEM AND THERE ARE NO FURTHER ISSUES. NO REPORTS OF PATIENT HARM WERE REPORTED DURING THIS EVENT. DURING THE INVESTIGATION, IT WAS DISCOVERED (B)(4) STUDY DATES WERE IMPACTED AND INCLUDED (B)(4) MRNS (MEDICAL RECORD NUMBERS). AGFA WILL SUBMIT A MDR TO THE FDA FOR EACH OF THE (B)(4) STUDY DATES AND INCLUDE THE SPECIFIC MRNS REFERENCING EACH STUDY DATE. THE MDR SUBMISSION NAMES FOR THE ENTIRE EVENT WILL INCLUDE: 1225058-2013-00001 THROUGH 1225058-2013-00185. STUDY DATE ON: (B)(6) 2011. MRNS: (B)(6).

Description of Event or Problem · 1

ON (B)(6) 2013, A LEAD ECHO SONOGRAPHER ALERTED AGFA THAT ULTRASOUND MEASUREMENTS WERE INSERTED IN THE WRONG LOCATION FOR 2 DIFFERENT PHILLIPS IE33 ULTRASOUND CARTS. THE CARDIOLOGIST WAS CONCERNED THAT ERRONEOUS MEASUREMENTS RESULTED IN INCORRECT DIAGNOSIS AS THE MITRAL REGURGITATION IS RATED HIGHER THAN IT ACTUALLY IS. ADDITIONALLY, THE SPECIFIC MEASUREMENT VALVES WERE NOT CONSISTENT WITH THE ACQUIRED IMAGE STUDIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
81111 IMPAX CV REPORTING IMPAX CV REPORTING LLZ AGFA HEALTHCARE CORP. IMPAX CV 7.4SU3 HF1 NA

Patients

Seq Age Sex Outcome Treatment
1