FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3000994 · Received February 27, 2013

Report

Report Number
1720753-2013-03176
Event Type
Malfunction
Date Received
February 27, 2013
Date of Event
February 18, 2013
Report Date
February 27, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND REPLACED BLOWN FUSES, THE CHARGER BOARD, AND THE CHARGER BOARD CABLES. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED AN A/D CHANNEL 8 FAILURE. THIS MEANS THE ANALOG-TO-DIGITAL CHANNEL FAILED DURING SYSTEM START-UP. THIS ERROR WILL LIKELY PREVENT THE SYSTEM FROM BOOTING UP. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
83536 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1