FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3000992 · Received February 27, 2013

Report

Report Number
1720753-2013-03180
Event Type
Malfunction
Date Received
February 27, 2013
Date of Event
February 11, 2013
Report Date
February 27, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND REPLACED THE SNUBBER BOARD, THE HIGH VOLTAGE SUPPLY REGULATOR, AND THE BATTERY CHARGER. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED AN ANNODE TEMPERATURE WARNING OF 65%. THE FIELD ENGINEER NOTED THAT THE SYSTEM WAS UNABLE TO PERFORM FLUOROSCOPY X-RAY. THERE IS NO REPOT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
83922 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1