FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3000971 · Received February 27, 2013

Report

Report Number
1720753-2013-03198
Event Type
Malfunction
Date Received
February 27, 2013
Date of Event
February 21, 2013
Report Date
February 27, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE PS1 POWER SUPPLY WAS REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A COMMUNICATION FAILURE ERROR MESSAGE AND LOCKED UP. THERE IS NO REPORT OF PT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
83924 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1