FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3000945 · Received February 27, 2013

Report

Report Number
1720753-2013-03191
Event Type
Malfunction
Date Received
February 27, 2013
Date of Event
February 7, 2013
Report Date
February 27, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN IN SITE INVESTIGATION. THE BACKPLANE, FLUORO FUNCTIONS BOARD, GENERATOR INTERFACE BOARD, SYSTEM INTERFACE BOARD, CPU, AND THE INTERCONNECT CABLE WERE REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THEY SYSTEM WOULD NOT BOOT UP. THERE WAS NO PT INJURY OR DEATH REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
83586 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1