FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 3000846 · Received March 13, 2013

Report

Report Number
3008382007-2013-04730
Event Type
Malfunction
Date Received
March 13, 2013
Report Date
February 12, 2013
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE METER AND TEST STRIPS INVOLVED WITH THIS COMPLAINT HAVE BEEN RETURNED BUT NOT YET EVALUATED BY LIFESCAN PRODUCT ANALYSIS. LFS WILL EVALUATE THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

(B)(4): THE REPORTED COMPLAINT WAS OBSERVED ON THE METER'S ERROR LOG; HOWEVER, PA WAS UNABLE TO REPRODUCE FAILURE IN METER TESTING. THE TEST STRIPS INVOLVED WITH THIS COMPLAINT WERE EVALUATED AND ER4 WAS OBSERVED WITH CONTROL SOLUTION TESTING. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING ERROR 4. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
105989 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3414372

Patients

Seq Age Sex Outcome Treatment
1 81 YR