FDA Adverse Event Malfunction Summary report: N

MICRO DRILL MEDIUM STRAIGHT ATTACHMENT

MDR report key: 3000737 · Received March 12, 2013

Report

Report Number
0001811755-2013-00512
Event Type
Malfunction
Date Received
March 12, 2013
Date of Event
February 20, 2013
Report Date
February 20, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER'S COMPLAINT WAS CONFIRMED; THE PROBABLE CAUSE OF THE DEVICE OVERHEATING WAS ATTRIBUTED TO CORRODED BEARINGS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A MANDIBULAR OSTEOTOMY, THE PATIENT RECEIVED A FIRST DEGREE BURN ON THE LIP. ADDITIONAL ANTIBIOTIC PRESCRIPTION WAS NOT GIVEN BECAUSE THE PATIENT RECEIVED AN ANTIBIOTIC PRIOR TO THE PROCEDURE. THE PATIENT WAS REPORTED AS DOING WELL ON THE FOLLOW UP VISIT. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
103996 MICRO DRILL MEDIUM STRAIGHT ATTACHMENT BONE CUTTING INSTRUMENT AND ACCESSORIES DZJ STRYKER INSTRUMENTS-KALAMAZOO 0308402953

Patients

Seq Age Sex Outcome Treatment
1 Other