FDA Adverse Event
Malfunction
Summary report: N
MICRO DRILL MEDIUM STRAIGHT ATTACHMENT
MDR report key: 3000737
·
Received March 12, 2013
Report
- Report Number
- 0001811755-2013-00512
- Event Type
- Malfunction
- Date Received
- March 12, 2013
- Date of Event
- February 20, 2013
- Report Date
- February 20, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE CUSTOMER'S COMPLAINT WAS CONFIRMED; THE PROBABLE CAUSE OF THE DEVICE OVERHEATING WAS ATTRIBUTED TO CORRODED BEARINGS.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A MANDIBULAR OSTEOTOMY, THE PATIENT RECEIVED A FIRST DEGREE BURN ON THE LIP. ADDITIONAL ANTIBIOTIC PRESCRIPTION WAS NOT GIVEN BECAUSE THE PATIENT RECEIVED AN ANTIBIOTIC PRIOR TO THE PROCEDURE. THE PATIENT WAS REPORTED AS DOING WELL ON THE FOLLOW UP VISIT. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 103996 | MICRO DRILL MEDIUM STRAIGHT ATTACHMENT | BONE CUTTING INSTRUMENT AND ACCESSORIES | DZJ | STRYKER INSTRUMENTS-KALAMAZOO | 0308402953 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |