FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3000648 · Received March 12, 2013

Report

Report Number
3004209178-2013-91795
Event Type
Malfunction
Date Received
March 12, 2013
Date of Event
February 25, 2013
Report Date
March 1, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ME
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING, IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. THE BLOOD GLUCOSE READING WAS 130MG/DL. TROUBLESHOOTING WAS PERFORMED. THE CUSTOMER WAS UNSURE IF THE DRIVE SUPPORT CAP WAS LOOSE, FLUSH OR PROTRUDED. THE DEVICE WAS NOT EXPOSED TO A HIGH MAGNETIC FIELD. ASSISTED THE CALLER TO RUN THE DISPLACEMENT TEST AND PASSED. ADVISED THE CUSTOMER TO DISCONTINUE THE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. DURING THE CALL THE CUSTOMER ALSO REPORTED BEING HOSPITALIZED DUE TO DIABETES KETOACIDOSIS AND HIGH BLOOD GLUCOSE OVER 600MG/DL. THE CUSTOMER EXPERIENCED VOMITING AND KETONES. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
103000 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAB

Patients

Seq Age Sex Outcome Treatment
1 9 YR