RESERVOIR 1.8ML
Report
- Report Number
- 3004209178-2013-91802
- Event Type
- Malfunction
- Date Received
- March 12, 2013
- Date of Event
- March 1, 2013
- Report Date
- March 2, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- FRN
- PMA / PMN Number
- K001828
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.
INSPECTED TWO OPENED AND USED RESERVOIRS. PERFORMED MANUAL PRIME AND HIGH PRESSURE TEST PER SPECIFICATIONS. RAN PRIMING IN INSULIN PUMP. RESERVOIR PASSED PER SPECIFICATION. NO LEAKAGE ANOMALIES WERE OBSERVED DURING ANALYSIS. CHECKED O-RINGS FOR DEFECTS AND NONE WERE FOUND. RESERVOIRS CONNECTED AND LOCKED IN PLACE PROPERLY.
THE CUSTOMER REPORTED THAT INSULIN FROM THE BOTTOM OF THE RESERVOIR LEAKED. THE BLOOD GLUCOSE READING WAS 150MG/DL. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 104197 | RESERVOIR 1.8ML | RESERVOIR | FRN | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-326A | H8486688 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 28 YR |