FDA Adverse Event Malfunction Summary report: N

MICRO OSCILLATING SAW WITH XL BLADE MOUNT

MDR report key: 3000436 · Received March 12, 2013

Report

Report Number
0001811755-2013-00505
Event Type
Malfunction
Date Received
March 12, 2013
Date of Event
February 13, 2013
Report Date
February 13, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
ERL
PMA / PMN Number
ERL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON DISASSEMBLY FOR VISUAL EXAMINATION, THE BEARINGS WERE FOUND TO BE GRITTY AND CORROSION WAS DISCOVERED INSIDE THE MOTOR. IN ADDITION, THE HOUSING AND ADAPTOR WERE WORN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE MICRO OSCILLATING SAW WITH XL BLADE MOUNT OVERHEATED AND RAN WITHOUT USER ACTIVATION DURING A PROCEDURE. THERE WERE NO PATIENT OR USER INJURIES, AND NO ADVERSE CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
103572 MICRO OSCILLATING SAW WITH XL BLADE MOUNT DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE ERL STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1