PROMUS ELEMENT ¿
Report
- Report Number
- 2134265-2013-01395
- Event Type
- Malfunction
- Date Received
- March 12, 2013
- Date of Event
- December 17, 2012
- Report Date
- February 18, 2013
- Manufacturer
- BOSTON SCIENTIFIC - GALWAY
- Product Code
- NIQ
- PMA / PMN Number
- SIMILAR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KR
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). DEVICE IS COMBINATION PRODUCT. DEVICE EVALUATED BY MFR: A VISUAL AND MICROSCOPIC EXAMINATION IDENTIFIED A KINK ON THE SHAFT AT THE LOCATION OF THE STENT AND THE BALLOON (13MM PROXIMAL TO THE TIP OF THE DEVICE). DUE TO THE KINK SOME STRUTS ON THE SECOND AND THE THIRD ROW ON THE DISTAL END OF THE STENT WERE RAISED UP. THIS DAMAGE IS CONSISTENT WITH THE STENT MEETING RESISTANCE UPON ADVANCEMENT AND/OR WITHDRAWAL. THE BALLOON AND THE TIP OF THE DEVICE WERE VISUALLY AND MICROSCOPICALLY EXAMINED AND NO ISSUES WERE NOTED WITH THEIR PROFILES THAT COULD HAVE POTENTIALLY CONTRIBUTED TO THE COMPLAINT INCIDENT. THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO ANY POSITIVE PRESSURE. NO KINKS OR DAMAGE WERE NOTICED ALONG THE SHAFT OF THE DEVICE. NO OTHER ISSUES WERE NOTED WITH THE DEVICE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).
THIS COMPLAINT IS NOW REPORTABLE BASED ON THE ANALYSIS COMPLETED ON (B)(6) 2013. IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION (PCI) THE DEVICE WAS UNABLE TO CROSS THE LESION. THE 78% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY LEFT CIRCUMFLEX CORONARY ARTERY (LCX). THE LESION WAS PREDILATED AND THEN A 2.50X16MM PROMUS ELEMENT STENT DELIVERY SYSTEM (SDS) WAS ADVANCED TO THE LCX; HOWEVER, THE DEVICE WAS UNABLE TO CROSS THE LESION. THE DEVICE WAS REMOVED FROM THE PATIENT AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S CURRENT CONDITION IS STABLE. HOWEVER, RETURNED PRODUCT ANALYSIS REVEALED STENT DAMAGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 103723 | PROMUS ELEMENT ¿ | STENT, CORONARY, DRUG-ELUTING | NIQ | BOSTON SCIENTIFIC - GALWAY | H7493911316250 | 14635390 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR |