PORTABLE VIDEO MONITOR, GVL-2000
MDR report key: 3000140
·
Received February 14, 2013
Report
- Report Number
- 9615393-2013-00063
- Event Type
- Malfunction
- Date Received
- February 14, 2013
- Date of Event
- January 16, 2013
- Report Date
- January 16, 2013
- Manufacturer
- VERATHON MEDICAL (CANADA) ULC.
- Product Code
- CCW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
SERVICE RECEIVED THE DEVICE FOR FURTHER EVAL. SERVICE PLUGGED IN THE MONITOR AND THE CHARGE LED WAS BLINKING. THEN SERVICE OPENED THE UNIT AND SAW THAT THE WIRE FROM THE SWITCH TO THE POWER PCBA (POWER CIRCUIT BOARD ASSEMBLY) WAS DISCONNECTED. SERVICE RESOLDERED THE WIRE AND PUT THE UNIT BACK TOGETHER. SERVICE CONFIRMED THE REPORTED FAILURE WHICH WAS DUE TO A LOOSE WIRE FROM THE SWITCH TO THE POWER PCBA.
Description of Event or Problem · 1
THE CUSTOMER CONTACTED VERATHON INC. REGARDING A PORTABLE VIDEO MONITOR, GVL-2000 THAT HAS A POWER MANAGEMENT ISSUE. THE PORTABLE VIDEO MONITOR, GVL-2000 WILL NOT HOLD CHARGE. THERE IS NO INJURY TO THE PT REPORTED IN THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 65386 | PORTABLE VIDEO MONITOR, GVL-2000 | NONE | CCW | VERATHON MEDICAL (CANADA) ULC. | 0231-0003 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|