FDA Adverse Event Malfunction Summary report: N

MEGASUTURECUT NEEDLEDRIVER INSTRUMENT

MDR report key: 3000115 · Received March 8, 2013

Report

Report Number
2955842-2013-00778
Event Type
Malfunction
Date Received
March 8, 2013
Date of Event
February 7, 2013
Report Date
February 7, 2013
Manufacturer
INTUITIVE SURGICAL, INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT ON GRIP CABLE WAS BROKEN AT THE DISTAL CLEVIS AND THERE WAS DAMAGE ON IDLER PULLEYS. THERE WERE MANY DENTS ALONG THE EDGES OF THE IDLER PULLEY, WHICH LIKELY CAUSED THE CABLE TO BREAK. ENGINEERING CONCLUDED THAT THE DENTS ON THE IDLER PULLEY WERE LIKELY CAUSED BY MISHANDLING. THE BROKEN CABLE SEGMENT STUCK OUT AT THE WRIST APPROXIMATELY .5155". NO OTHER DAMAGE WAS FOUND. THE INSTRUMENT AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS. HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SI SURGICAL PROCEDURE, THE CABLE ON THE MEGASUTURECUT NEEDLEDRIVER INSTRUMENT WAS FOUND TO BE BROKEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
98780 MEGASUTURECUT NEEDLEDRIVER INSTRUMENT ENDOSCOPIC INSTRUMENT, NAY NAY INTUITIVE SURGICAL, INC. 420309-01 M10121112 732

Patients

Seq Age Sex Outcome Treatment
1 ACCESSORIES| DA VINCI SI SURG. SYS.| INSTRUMENTS