FDA Adverse Event Malfunction Summary report: N

PK DISSECTING FORCEPS INSTRUMENT

MDR report key: 3000097 · Received March 8, 2013

Report

Report Number
2955842-2013-00780
Event Type
Malfunction
Date Received
March 8, 2013
Date of Event
February 7, 2013
Report Date
February 7, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE INSTRUMENT WAS RETURNED WITH A FRAYED PITCH LOOSE CABLE AT DISTAL CLEVIS HUB. NO DAMAGE WAS FOUND AT THE CLEVIS AND NO OTHER CABLE DAMAGE WAS FOUND. THE INSTRUMENT PASSED ELECTRICAL CONTINUITY TESTING. ENGINEERING EVALUATION ALSO FOUND THAT ONE GRIP WAS BENT, CAUSING SIDE TO SIDE MISALIGNMENT OF GRIPS. THERE WAS A .038 OFFSET AT THE TIPS, INDICATING OVERLOADING AT THE TIP. ENGINEERING CONCLUDED THAT THE DAMAGE WAS LIKELY DUE TO MISHANDLING. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS,, O HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS.

Additional Manufacturer Narrative · 1

THE INSTRUMENT HAS NOT BEEN RETURNED FOR EVALUATION, THEREFORE, THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE CANNOT BE DETERMINED. A FOLLOW-UP MDR WILL BE SUBMITTED IF THE INSTRUMENT IS RETURNED (POST ENGINEERING EVALUATION) OR IF ADDITIONAL INFORMATION IS RECEIVED. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING SET UP OF A DA VINCI SI SURGICAL PROCEDURE THE WIRE AT THE DISTAL END OF THE PK DISSECTING FORCEPS INSTRUMENTS WAS NOTED TO BE FRAYED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
98739 PK DISSECTING FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT, NAY NAY INTUITIVE SURGICAL,INC. 420227-03 M10120621 483

Patients

Seq Age Sex Outcome Treatment
1 ESU| INSTRUMENTS| DA VINCI SI SURG. SYS.| ACCSS