Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 3000065
·
Received March 8, 2013
Report
- Report Number
- 8020893-2013-00573
- Event Type
- Malfunction
- Date Received
- March 8, 2013
- Date of Event
- February 13, 2013
- Report Date
- February 13, 2013
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- RESPIRATORY THERAPIST
Narratives
Description of Event or Problem · 1
THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) REPORTED THAT AN 840 VENTILATOR HAD AN UNREADABLE DISPLAY. NO PATIENT INVOLVEMENT. THE CSE INSPECTED THE DEVICE AND REPLACED THE GRAPHIC USER INTERFACE (GUI) PCG. THE UNIT PASSED EXTENDED SELF-TESTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 98915 | 840 VENTILATOR | VENTILATOR | CBK | COVIDIEN, FORMERLY NELLCOR | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|