UNIV 2-HOLE SHL 56MM LNR SZ 24
Report
- Report Number
- 0001825034-2013-00577
- Event Type
- Injury
- Date Received
- March 12, 2013
- Date of Event
- January 28, 2013
- Report Date
- February 14, 2013
- Manufacturer
- BIOMET ORTHOPEDICS
- Product Code
- LPH
- PMA / PMN Number
- PK030055
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- PHYSICIAN
Narratives
DIMENSIONAL EVALUATION FOUND COMPONENT TO BE WITHIN APPROPRIATE DESIGN SPECIFICATION. EXAMINATION OF RETURNED DEVICE WAS INCONCLUSIVE. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, NUMBER 4 STATES, "LOOSENING OR MIGRATION OF THE IMPLANTS CAN OCCUR DUE TO LOSS OF FIXATION, TRAUMA, MALALIGNMENT, BONE RESORPTION, OR EXCESSIVE ACTIVITY."
EVALUATION IN PROCESS BUT NOT YET COMPLETE. UPON COMPLETION OF EVALUATION, A FOLLOW UP REPORT WILL BE SENT TO THE FDA. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, NUMBER 4 STATES, "LOOSENING OR MIGRATION OF THE IMPLANTS CAN OCCUR DUE TO LOSS OF FIXATION, TRAUMA, MALALIGNMENT, BONE RESORPTION, OR EXCESSIVE ACTIVITY."
IT WAS REPORTED PATIENT UNDERWENT TOTAL HIP ARTHROPLASTY IN (B)(6) 2010. SUBSEQUENTLY, PATIENT WAS REVISED ON (B)(6) 2013 DUE TO CUP LOOSENING AND PAIN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 103074 | UNIV 2-HOLE SHL 56MM LNR SZ 24 | PROSTHESIS, HIP | LPH | BIOMET ORTHOPEDICS | N/A | 687020 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Hospitalization| R |