FDA Adverse Event Malfunction Summary report: N

HS III PROXIMAL SEAL

MDR report key: 3000027 · Received March 8, 2013

Report

Report Number
2242352-2013-00215
Event Type
Malfunction
Date Received
March 8, 2013
Date of Event
February 5, 2013
Report Date
February 14, 2013
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
DXC
PMA / PMN Number
K080169
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION RESULTS: THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. IT SHOWED NO SIGNS OF CLINICAL USAGE OR EVIDENCE OF BLOOD. THE DELIVERY DEVICE WAS RETURNED OUTSIDE OF THE LOADING DEVICE. THE TENSION SPRING ASSEMBLY, ANCHOR TAB AND THE SEAL WERE LOCATED INSIDE THE BODY OF THE LOADING DEVICE. THE GREEN SLIDE LOCK REMAINED LOCKED; THE WHITE PLUNGER WAS NOT DEPRESSED ON THE DELIVERY DEVICE. BASED UPON THE EVALUATION RESULTS, THE REPORTED COMPLAINT WAS CONFIRMED FOR FAILURE TO LOAD. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCE RECORDED IN THE LOT HISTORY. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, THE HEARTSTRING III SEAL REMAINED INSIDE OF THE LOADING UPON REMOVAL FROM THE DELIVERY DEVICE. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
98820 HS III PROXIMAL SEAL CLAMPLESS BEATING HEART DXC MAQUET CARDIOVASCULAR, LLC HS-3045 25068871

Patients

Seq Age Sex Outcome Treatment
1 NA