FDA Adverse Event Malfunction Summary report: N

VASOVIEW HEMOPRO 2

MDR report key: 3000011 · Received March 8, 2013

Report

Report Number
2242352-2013-00220
Event Type
Malfunction
Date Received
March 8, 2013
Date of Event
February 13, 2013
Report Date
February 14, 2013
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
GEI
PMA / PMN Number
K101274
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT YET BEEN RETURNED TO MAQUET CARDIAC SURGERY FOR EVALUATION. WE ARE FOLLOWING UP WITH THE CUSTOMER FOR THE RETURN OF THE DEVICE. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF THE DEVICE IS RECEIVED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCE RECORDED IN THE LOT HISTORY. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE USING THE HEMOPRO DEVICE, THE C-RING FAILED TO FULLY RETRACT INTO THE CANNULA. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PT EFFECTS. THE HOSPITAL INTENDS TO RETURN THE PRODUCT IN QUESTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
100359 VASOVIEW HEMOPRO 2 ENDOSCOPIC VESSEL HARVESTING GEI MAQUET CARDIOVASCULAR, LLC VH-4000 25067691

Patients

Seq Age Sex Outcome Treatment
1 NA