FDA Adverse Event Malfunction Summary report: N

5.5MM TI MIN INVASIVE CURVED ROD 200MM RADIUS/125MM

MDR report key: 2999364 · Received March 12, 2013

Report

Report Number
8030965-2013-00830
Event Type
Malfunction
Date Received
March 12, 2013
Date of Event
February 14, 2013
Report Date
February 14, 2013
Manufacturer
SYNTHES GMBH
Product Code
NKB
PMA / PMN Number
K093668
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

510K#: K093668. THE PERFORMED INVESTIGATION HAS SHOWN THAT THE MACHINED END OF THE MIS ROD HAS A NOTCH WHICH DOES NOT MEET THE SPECIFICATIONS. HOWEVER, THIS DOES NOT AFFECT THE FUNCTION OF THE MIS ROD WHICH EASILY CAN BE CONNECTED TO THE ROD HOLDER. THE COMPLAINED MIS ROD WAS SENT FOR FURTHER INVESTIGATION TO THE MANUFACTURING PLANT. THE RESULTS SHOW THAT THIS IMPLANT HAS BEEN PRODUCED IN MAY 2011 AND NO ABNORMALITIES WERE FOUND ON PRODUCTION DOCUMENTS. NO FURTHER COMPLAINT ON THIS ITEM IS KNOWN. BASED ON THESE RESULTS, NO FURTHER MEASURES ARE INITIATED AND THIS CASE IS EVALUATED AS AN ISOLATED CASE. THIS DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. PLACEHOLDER.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. ADDITIONAL PRODUCT CODES: MNH, MNI, KWO, KWP. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS THE PRODUCT IS ENTERING THE COMPLAINT SYSTEM. THE DEVICE HISTORY RECORD REVIEW REPORTED THE FOLLOWING: MANUFACTURING DOCUMENTS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: DURING THE REFILL OF A LOAN SET IT WAS NOTICED THAT THERE WAS A DEFORMITY IN THE CONNECTION PART OF THE MIS ROD. THIS IS 1 OF 1 REPORT FOR THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
102824 5.5MM TI MIN INVASIVE CURVED ROD 200MM RADIUS/125MM NKB SYNTHES GMBH 3802893

Patients

Seq Age Sex Outcome Treatment
1