5.5MM TI MIN INVASIVE CURVED ROD 200MM RADIUS/125MM
Report
- Report Number
- 8030965-2013-00830
- Event Type
- Malfunction
- Date Received
- March 12, 2013
- Date of Event
- February 14, 2013
- Report Date
- February 14, 2013
- Manufacturer
- SYNTHES GMBH
- Product Code
- NKB
- PMA / PMN Number
- K093668
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER
Narratives
510K#: K093668. THE PERFORMED INVESTIGATION HAS SHOWN THAT THE MACHINED END OF THE MIS ROD HAS A NOTCH WHICH DOES NOT MEET THE SPECIFICATIONS. HOWEVER, THIS DOES NOT AFFECT THE FUNCTION OF THE MIS ROD WHICH EASILY CAN BE CONNECTED TO THE ROD HOLDER. THE COMPLAINED MIS ROD WAS SENT FOR FURTHER INVESTIGATION TO THE MANUFACTURING PLANT. THE RESULTS SHOW THAT THIS IMPLANT HAS BEEN PRODUCED IN MAY 2011 AND NO ABNORMALITIES WERE FOUND ON PRODUCTION DOCUMENTS. NO FURTHER COMPLAINT ON THIS ITEM IS KNOWN. BASED ON THESE RESULTS, NO FURTHER MEASURES ARE INITIATED AND THIS CASE IS EVALUATED AS AN ISOLATED CASE. THIS DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. PLACEHOLDER.
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. ADDITIONAL PRODUCT CODES: MNH, MNI, KWO, KWP. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS THE PRODUCT IS ENTERING THE COMPLAINT SYSTEM. THE DEVICE HISTORY RECORD REVIEW REPORTED THE FOLLOWING: MANUFACTURING DOCUMENTS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND.
DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: DURING THE REFILL OF A LOAN SET IT WAS NOTICED THAT THERE WAS A DEFORMITY IN THE CONNECTION PART OF THE MIS ROD. THIS IS 1 OF 1 REPORT FOR THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 102824 | 5.5MM TI MIN INVASIVE CURVED ROD 200MM RADIUS/125MM | NKB | SYNTHES GMBH | 3802893 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |