FDA Adverse Event Death Summary report: N

AGILENT TECHNOLOGIES/VIRIDIA

MDR report key: 299763 · Received October 5, 2000

Report

Report Number
299763
Event Type
Death
Date Received
October 5, 2000
Date of Event
May 28, 2000
Report Date
October 2, 2000
Manufacturer
AGILENT TECHNOLOGIES INC
Product Code
DRT
Product Problem
Yes
Report Source
User Facility report
Reporter Location
MA, US
Reporter Occupation
INVALID DATA

Narratives

Description of Event or Problem · 1

THE PT WENT INTO AN EXTENDED EPISODE OF VENTRICULAR FIBRILLATION WHILE ON CARDIAC MONITORING WITH TELEMETRY. THE AUDIBLE ALARM (A SPEAKER) AT THE CENTRAL MONITOR FAILED TO ANNUNCIATE, AND AS A RESULT, THE EVENT WENT UNNOTICED BY CLINICAL PERSONNEL, AND THE VENTRICULAR FIBRILLATION PROGRESSED TO AN UNATTENDED CARDIOPULMONARY ARREST. THE PAPER RECORDING THAT WAS GENERATED AT THE TIME OF THE EVENT BY THE MONITOR CLEARLY DOCUMENTS THE SERIES OF EVENTS LEADING TO THE ARREST. EXAMINATION OF THE MONITORING SYSTEM REVEALED THE FOLLOWING: THE CONNECTOR FROM THE ALARM SPEAKER TO THE COMPUTER GENERATING THE ALARM WAS A SIMPLE "BANANA PLUG" WITH NO MEANS FOR FIXATION OR SECURE CONNECTION TO THE BODY OF THE COMPUTER. THE INITIAL INSTALLATION OF THE MONITOR BY TECHNICIANS FROM AGILENT TECHNOLOGIES IN DECEMBER OF 1999 WAS NOT DONE IN ACCORDANCE WITH AGILENT'S WRITTEN PROCEDURE (SEE ATTACHMENT B) WHICH CALLS FOR THE SPEAKER WIRE TO BE ATTACHED TO A "SECURE" CONNECTOR, I.E., ONE THAT IS FIXED TO THE BODY OF THE MONITOR WITH A THREADED ONNECTION, WITH A ELECTRICAL TIE WRAP. THIS PROCEDURE, EVEN WHEN PERFORMED ACCORDING TO THE MFR'S INSTRUCTIONS, FAILS TO SECURE THE SPEAKER CONNECTOR, THE KEYBOARD CONNECTOR, AND THE MOUSE CONNECTOR SUFFICIENTLY TO PREVENT THESE DEVICES FROM BECOMING DISCONNECTED. INADVERTENT DISCONNECTION OF THE MOUSE OR THE KEYBOARD MAY CAUSE THE COMPUTER TO "FREEZE" AND THEREBY FAIL TO IDENTIFY AND ANNUNCIATE WHEN AN ARRYTHMIA OCCURS. FURTHER EXAMINATION OF THE SYSTEM REVEALED THAT AT THE TIME THE SYSTEM WAS ASSEMBLED AND INSTALLED BY AGILENT TECHNOLOGIES, CRITICAL COMPONENTS, E.G. THE COMPUTER AND MONITOR FOR THE SYSTEM, WERE PLUGGED INTO THE "SURGE CONTROL" SIDE OF THE UNINTERRUPTABLE POWER SUPPLY, RATHER THAN THE UPS POWER SIDE, AND WOULD HAVE FAILED IMMEDIATELY IN THE EVENT OF A POWER OUTAGE. TESTING OF THE UPS REVEALED THAT WITH THE COMPUTER AND MONITOR CONNECTED TO THE UPS AND THE PRINTER DISCONNECTED (MINIMUM OPERABLE CONFIGURATION) THE UPS PROVIDED ONLY FIVE MINUTES OF BACK-UP POWER. IT IS FACILITY'S OPINION THAT THIS LENGTH OF TIME IS AN INADEQUATE SAFE INTERVAL FOR CLINICIANS TO RESPOND TO SUCH AN OUTAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AGILENT TECHNOLOGIES/VIRIDIA CARDIAC MONITOR/ALARM DRT AGILENT TECHNOLOGIES INC M3150A *

Patients

Seq Age Sex Outcome Treatment
1 83 YR Death