FDA Adverse Event
Malfunction
Summary report: N
ORTHO SUMMIT SAMPLE HANDLING SYSTEM
MDR report key: 2994319
·
Received March 8, 2013
Report
- Report Number
- 2250051-2013-00066
- Event Type
- Malfunction
- Date Received
- March 8, 2013
- Date of Event
- February 8, 2013
- Report Date
- March 8, 2013
- Manufacturer
- HAMILTON BONADUZ AG
- Product Code
- MMH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ON (B)(4) 2013 AN OCD (FE) ARRIVED AT THE CUSTOMER SITE AND INSPECTED ALL TIP AND PLUNGER CLAMPS. THE FE INSPECTED ALL EJECTION PINS. THE FE TESTED THE LLD WITH SPLLD TEST. THE FE TESTED THE SUMMIT WITH THE OAS USER SOFTWARE TEST. ALL TESTS PASSED. REPAIRS HAVE RETURNED THIS INSTRUMENT TO EXPECTED OPERATION.
Description of Event or Problem · 1
THE ORTHO SUMMIT SAMPLE HANDLING SYSTEM INSTRUMENT DID NOT PIPETTE THE CORRECT AMOUNT OF SAMPLE AND DID NOT GENERATE AN ERROR MESSAGE. NO DEATH OR SERIOUS INJURY WAS ASSOCIATED WITH THIS INCIDENT. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 98607 | ORTHO SUMMIT SAMPLE HANDLING SYSTEM | SAMPLE PIPETTOR | MMH | HAMILTON BONADUZ AG | 936400 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |