FDA Adverse Event Malfunction Summary report: N

QUICK COUPLING F/DRILL BITS FORTRS

MDR report key: 2994230 · Received March 8, 2013

Report

Report Number
2520274-2013-01326
Event Type
Malfunction
Date Received
March 8, 2013
Date of Event
January 21, 2013
Report Date
February 9, 2013
Manufacturer
SYNTHES (USA)
Product Code
HWE
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE IS USED FOR TREATMENT, NOT DIAGNOSIS. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS.

Additional Manufacturer Narrative · 1

WE DETECTED THAT GREASE WAS LEAKING OUT OF THE REFERENCED ARTICLE. POSSIBLE CAUSES: TOO MUCH GREASE FILLED IN TO THE ARTICLES, THE SEALING COULD BE DEFECTIVE, WRONG PURIFICATION.

Description of Event or Problem · 1

A DEVICE REPORT FROM SYNTHES (B)(4) PROVIDES INFORMATION FROM A FACILITY IN (B)(6) REGARDING DEVICES ASSOCIATED WITH THE TRAUMA RECON SYSTEM. PRIOR TO A PROCEDURE TO TREAT A PATIENT WITH A FRACTURE OF THE SHAFT OF THE LOWER LEG, A NURSE BROKE THE SEAL OF THE STERILIZED PACKAGE AND FOUND NOTICEABLE IMPURITIES, DESCRIBED AS A GRAY-COLOURED OILY SUBSTANCE, ADHERED TO THE SURFACE OF THE TRS DEVICE. THE NURSE WIPED OUT THE IMPURITIES ON THE SURFACE OF THE TRS DEVICE AND THE DOCTOR PROCEEDED TO USE THE DEVICE. THIS IS REPORT NUMBER 2 OF 7 FOR THE SAME EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
99292 QUICK COUPLING F/DRILL BITS FORTRS HWE SYNTHES (USA)

Patients

Seq Age Sex Outcome Treatment
1