FDA Adverse Event Malfunction Summary report: N

IMPL-HOLDER F/SYNFIX-LR

MDR report key: 2994205 · Received March 8, 2013

Report

Report Number
8030965-2013-00787
Event Type
Malfunction
Date Received
March 8, 2013
Date of Event
February 7, 2013
Report Date
February 11, 2013
Manufacturer
SYNTHES GMBH
Product Code
LXH
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS USED FOR TREATMENT, NOT DIAGNOSIS. THE INVESTIGATIONS HAVE SHOWN THAT THE TIP IS INDEED BROKEN OFF. UNFORTUNATELY WE ARE NOT ABLE TO COMPREHEND HOW THIS BREAKAGE OCCURRED. WE CAN ONLY SUPPOSE THOUGH THAT THE SCREWDRIVER HAD NOT BEEN SCREWED IN DEEP ENOUGH INTO THE CAGE AND THAT THE TIP BROKE OFF DURING HAMMERING. OR THE SURGEON APPLIED TOO MUCH MECHANICAL FORCE AS HE EXPERIENCED DIFFICULTY UNCOUPLING IT FROM THE IMPLANT. THE BROKEN SURFACE IS HOMOGENEOUS WHICH INDICATES MATERIAL CONFORMITY AS WELL. THE MANUFACTURING RECORDS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND. NO PRODUCT FAULT COULD BE DETECTED.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS THE PRODUCT IS ENTERING THE COMPLAINT SYSTEM. THE DEVICE HISTORY RECORD REVIEW REPORTED THE FOLLOWING: MANUFACTURING DOCUMENTS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND. PLACEHOLDER.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES (B)(6) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: DURING A PROCEDURE ON (B)(6) 2013, DURING INSERTION OF THE SYNFIX CAGE THE THREADED TIP OF THE SYNFIX IMPLANT HOLDER BROKE OFF IN-SITU WHILE ATTACHED TO THE IMPLANT. REPORTEDLY THE SURGEON FELT DIFFICULTY UNCOUPLING THE INSTRUMENT AND THE IMPLANT. THIS IS 1 OF 1 REPORT FOR THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
98949 IMPL-HOLDER F/SYNFIX-LR LXH SYNTHES GMBH 2701876

Patients

Seq Age Sex Outcome Treatment
1 41 YR