FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2994161 · Received February 13, 2013

Report

Report Number
1720753-2013-01737
Event Type
Malfunction
Date Received
February 13, 2013
Date of Event
January 31, 2013
Report Date
February 13, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SVC REP PERFORMED AN ONSITE INVESTIGATION. THE IGBT SNUBBER WAS REPLACED, THE HIGH VOLTAGE SUPPLY REGULATOR WAS ADJUSTED AND A FILAMENT CALIBRATION WAS PERFORMED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYS HAD NO VIDEO. THIS RESULTED IN A LOSS OF THE LIVE IMAGE. THERE IS NO REPORT OF PT INJURY.