ILS 29MM, CURVED
Report
- Report Number
- 3005075853-2013-01120
- Event Type
- Malfunction
- Date Received
- March 8, 2013
- Report Date
- February 15, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K983536
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE ARRIVED WITH THE ANVIL MISSING, OTHERWISE IN GOOD VISUAL CONDITION. AS THE ORIGINAL ANVIL WAS NOT RECEIVED, FURTHER INVESTIGATION WITH THE ORIGINAL ANVIL COULD NOT BE PERFORMED. THE BREAKAWAY WASHER WAS NOT PRESENT AND THERE WERE NO STAPLES PRESENT. THE DEVICE WAS RELOADED WITH STAPLES, A NEW WASHER WAS PLACED ON THE DEVICE AND IT WAS TESTED FOR FUNCTIONALITY WITH A TEST ANVIL; IT FIRED AND FORMED ALL THE STAPLES AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE MANUFACTURING RECORDS WERE REVIEWED AND NO DISCREPANCIES WERE FOUND DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING LAPAROSCOPIC SIGMA PROCEDURE, THERE WAS AN INCOMPLETE STAPLE LINE. THEY SUTURED BY HAND TO COMPLETE THE PROCEDURE. NO FURTHER INFORMATION IS AVAILABLE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 100246 | ILS 29MM, CURVED | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |