FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2994149 · Received February 13, 2013

Report

Report Number
1720753-2013-01719
Event Type
Malfunction
Date Received
February 13, 2013
Date of Event
February 1, 2013
Report Date
February 13, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REP EVALUATED THE SYSTEM AND CLEANED AND LUBRICATED THE HIGH VOLTAGE CANDLESTICK CONNECTORS. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM SHUT ITSELF OFF AND REBOOTED. THERE IS NO REPORT OF PT INJURY OR DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
64031 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1