FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 2994143 · Received March 8, 2013

Report

Report Number
2531779-2013-02500
Event Type
Malfunction
Date Received
March 8, 2013
Report Date
February 11, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).

Additional Manufacturer Narrative · 1

FOLLOW UP #1 (B)(4) DEVICE EVALUATION: THE INSULIN PUMP HAS BEEN RETURNED AND ADDITIONAL INVESTIGATION WAS PERFORMED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: ON EXAMINATION, THERE WAS NO VISIBLE DAMAGE TO THE KEYPAD. ON TESTING, THE OK AND CONTRAST BUTTONS HAD INTERMITTENT RESPONSE AND REQUIRED MULTIPLE PRESSES WITH INCREASED FORCE TO EVOKE A RESPONSE. THE REMAINDER OF THE BUTTONS HAD NORMAL RESPONSE. THE KEYPAD COVER WAS REMOVED FOR INVESTIGATION AND REVEALED CONTAMINATION UNDER THE CONTACTS OF ALL THE BUTTONS.

Description of Event or Problem · 1

ON (B)(6) 2013, THE DISTRIBUTOR CONTACTED ANIMAS STATING THAT THE UP/DOWN ARROW AND OK KEYPAD BUTTONS WERE UNRESPONSIVE. THERE WAS NO REPORTED ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THERE WAS NO ADDITIONAL INFORMATION AVAILABLE FOR THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED DUE TO THE ALLEGED KEYPAD ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
100117 ANIMAS VIBE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1