FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2994134 · Received February 13, 2013

Report

Report Number
1720753-2013-01715
Event Type
Malfunction
Date Received
February 13, 2013
Date of Event
January 28, 2013
Report Date
February 13, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REP EVALUATED THE SYSTEM AND REPLACED THE GENERATOR INTERFACE BOARD AND DISCUSSED POWER LINE MONITORING WITH THE CUSTOMER. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYSTEM WOULD NOT COMPLETE BOOT UP. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
63815 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1