FDA Adverse Event Malfunction Summary report: N

COYOTE¿ ES

MDR report key: 2994119 · Received March 8, 2013

Report

Report Number
2134265-2013-01405
Event Type
Malfunction
Date Received
March 8, 2013
Date of Event
February 13, 2013
Report Date
February 13, 2013
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LIT
PMA / PMN Number
K080982
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATED BY MFR.: THE DEVICE HAS NOT BEEN RECEIVED FOR ANALYSIS. UPON RECEIPT AND COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE A BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED LEFT PERONEAL ARTERY. THE 3.0 X 20MM COYOTE ES OTW BALLOON WAS INFLATED AND RUPTURED ON THE 1ST INFLATION BELOW 14ATMS. THE DEVICE WAS REMOVED INTACT. THE PROCEDURE WAS COMPLETED WITH A NON BSC BALLOON. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS OK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
100023 COYOTE¿ ES CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL LIT BOSTON SCIENTIFIC - MAPLE GROVE H74939134302010 14894885

Patients

Seq Age Sex Outcome Treatment
1 82 YR