ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2013-02494
- Event Type
- Malfunction
- Date Received
- March 8, 2013
- Report Date
- February 8, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP HISTORY REVEALED MULTIPLE OCCLUSION ALARMS. INVESTIGATION REVEALED A "NO CARTRIDGE DETECTED" WARNING WAS EMITTED DURING THE LOAD STEP. THE PUMP WAS OPENED, AND A COMPONENT ON THE PRINT CIRCUIT BOARD (PCB) WAS FOUND TO BE MISALIGNED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT A COMPONENT ON THE PRINT CIRCUIT BOARD (PCB) WAS FOUND TO BE MISALIGNED. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2013.