FDA Adverse Event Malfunction Summary report: N

11MM/130 DEG TI CANN TROCH FIXATION NAIL 320MM/RIGHT-STER

MDR report key: 2993976 · Received March 8, 2013

Report

Report Number
1719045-2013-10306
Event Type
Malfunction
Date Received
March 8, 2013
Date of Event
January 26, 2012
Report Date
January 26, 2012
Manufacturer
SYNTHES MONUMENT
Product Code
HSB
PMA / PMN Number
K011857
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO FDA WARNING LETTER DATED FEBRUARY 2012. BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN, UNAVAILABLE OR UNCHANGED. DEVICE(S) LISTED IN THIS REPORT IS (ARE) USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR.(B)(4)

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO FDA WARNING LETTER DATED FEBRUARY 2012. DEVICE(S) LISTED IN THIS REPORT IS (ARE) USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. A REVIEW OF THE DEVICE HISTORY RECORDS WAS PERFORMED AND NO COMPLAINT RELATED ISSUES WERE FOUND. THE PRODUCT DEVELOPMENT EVALUATION REVEALED BASED ON EXAMINATION OF THE RETURNED PARTS, THE LOCKING MECHANISM WAS IN THE LOCKED DOWN POSITION WHEN ATTEMPTING TO INSERT THE HELICAL BLADE. THE LOCKING TAB WAS SEVERELY BENT AND BROKE OFF AS A RESULT. THIS COMPLAINT IS VALID, AND AN INTERNAL CORRECTIVE ACTION HAS BEEN OPENED TO ADDRESS THIS ISSUE. ADDITIONAL PRODUCT CODES - HWC.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A TROCHANTERIC FIXATION NAIL (TFN) PROCEDURE THE HELICAL BLADE WOULD ONLY ADVANCE THREE FOURTHS OF THE WAY DOWN, AND THE BLADE GOT STUCK IN THE NAIL. THE SURGEON INSERTED THE NAIL IN PATIENT, THE LOCKING MECHANISM ON BRAND NEW NAIL WAS DOWN AND DEPLOYED WHEN OPENED. SURGEON BACKED UP THE LOCKING MECHANISM ON THE NAIL A LITTLE UNTIL IT STOPPED, HE THEN INSERTED THE NAIL. IT WAS THEN WHEN HE WENT TO INSERT THE HELICAL BLADE AND THE BLADE WOULD ONLY ADVANCE ABOUT PART OF THE WAY AND THE BLADE GOT STUCK IN THE NAIL. SURGEON USED THE EXTRACTOR TO BACK OUT THE BLADE AND HE REMOVED THE NAIL. SURGEON USED A BRAND NEW NAIL AND BLADE TO COMPLETE THE PROCEDURE WITH NO FURTHER PROBLEMS. SURGERY WAS EXTENDED BY TEN TO FIFTEEN MINUTES, WITH NO HARM TO THE PATIENT. THIS IS REPORT 1 OF 2 FOR THIS EVENT.

Description of Event or Problem · 1

THIS IS REPORT 1 OF 2 FOR THIS COMPLAINT (B)(4)

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
99602 11MM/130 DEG TI CANN TROCH FIXATION NAIL 320MM/RIGHT-STER HSB SYNTHES MONUMENT 6209219

Patients

Seq Age Sex Outcome Treatment
1