FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 2993957 · Received March 1, 2013

Report

Report Number
1627487-2013-10068
Event Type
Injury
Date Received
March 1, 2013
Date of Event
February 5, 2013
Report Date
February 5, 2013
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS CHARGER MODEL NUMBER WAS ASSOCIATED IN A FIELD CORRECTION. MFR'S EVAL: CORRECTIVE AND PREVENTIVE ACTION (CAPA). RESULTS: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2 REFERENCE MFR REPORT: 1627487-2013-10067. IT WAS REPORTED THE PATIENT ((B)(6)) EXPERIENCES HEATING WHILE CHARGING HIS IPG STATING THAT THE SKIN AROUND THE IPG SITE GETS VERY HOT. IT WAS ALSO REPORTED THE PATIENT¿S RECHARGE BURDEN HAS INCREASED. A REPLACEMENT CHARGING SYSTEM WAS ISSUED TO THE PATIENT. IN ADDITION, THE PATIENT WAS PROVIDED WITH THE MFR SUGGESTIONS TO AVOID HEAT WHILE CHARGING THE IPG. FOLLOW UP INFO REVEALED THE REPLACEMENT CHARGING SYSTEM HAS IMPROVED THE PATIENT¿S RECHARGE BURDEN; HOWEVER, HE IS STILL EXPERIENCING THE HEATING SENSATION WHEN CHARGING HIS IPG. IT WAS REPORTED THE NEXT PLANNED COURSE OF ACTION WILL BE TO REPROGRAM THE IPG. PLEASE NOTE: THE IPG WAS IMPLANTED IN 2009 (EXACT DATE UNK).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
89256 CHARGING SYSTEM SCS CHARGING SYSTEM LGW ST. JUDE MEDICAL - NEUROMODULATION 3711 3825913

Patients

Seq Age Sex Outcome Treatment
1 71 YR Required Intervention