CHARGING SYSTEM
Report
- Report Number
- 1627487-2013-10068
- Event Type
- Injury
- Date Received
- March 1, 2013
- Date of Event
- February 5, 2013
- Report Date
- February 5, 2013
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS CHARGER MODEL NUMBER WAS ASSOCIATED IN A FIELD CORRECTION. MFR'S EVAL: CORRECTIVE AND PREVENTIVE ACTION (CAPA). RESULTS: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 2 OF 2 REFERENCE MFR REPORT: 1627487-2013-10067. IT WAS REPORTED THE PATIENT ((B)(6)) EXPERIENCES HEATING WHILE CHARGING HIS IPG STATING THAT THE SKIN AROUND THE IPG SITE GETS VERY HOT. IT WAS ALSO REPORTED THE PATIENT¿S RECHARGE BURDEN HAS INCREASED. A REPLACEMENT CHARGING SYSTEM WAS ISSUED TO THE PATIENT. IN ADDITION, THE PATIENT WAS PROVIDED WITH THE MFR SUGGESTIONS TO AVOID HEAT WHILE CHARGING THE IPG. FOLLOW UP INFO REVEALED THE REPLACEMENT CHARGING SYSTEM HAS IMPROVED THE PATIENT¿S RECHARGE BURDEN; HOWEVER, HE IS STILL EXPERIENCING THE HEATING SENSATION WHEN CHARGING HIS IPG. IT WAS REPORTED THE NEXT PLANNED COURSE OF ACTION WILL BE TO REPROGRAM THE IPG. PLEASE NOTE: THE IPG WAS IMPLANTED IN 2009 (EXACT DATE UNK).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 89256 | CHARGING SYSTEM | SCS CHARGING SYSTEM | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3711 | 3825913 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR | Required Intervention |